Tisch: TE-6070

High Volume PM10 Mass Flow controlled (MFC) Monitor

in Ambient Air

 

Complete with Size selective inlet, outdoor anodized aluminum shelter, aluminum blower motor assembly, 8" x 10" stainless steel filter holder, mass flow controller with 20/60 scfm flow probe,7-day mechanical timer, elapsed time indicator, continuous flow/pressure recorder, and filter paper cartridge (110v/60hz or 220v/50hz)

Specifications

 

Inlet collection efficiency:

Meets EPA’s PM-10 Code of Federal Regulations Appendix J to Part 50

Flow rate:

40 cfm (adjustable) (1.02 to 1.24m³)

Filter media:

Micro-Quartz filter media 8" x 10" for PM10 (TE-QMA)

Flow control:

MFC (TE-300-310)

Motor blower:

2-stage vacuum 0.6 HP 110v/60hz (TE-116311), 220v/50hz (TE-116312)

Flow Indicator:

Flow/pressure recorder (TE-5009) with TE-106 Charts

Mechanical programmable timer

7-day mechanical timer (TE-5007)

Elapsed time indicator device:

Elapsed Time Indicator (TE-5012)

Sample Module:

Particulate/vapor sampling module with Tischaloy coating (TE-1002)

    

Shipping Information

 

Size:

22.5" x 20" x 45.5" sampler carton, 32" x 32" x 25.5" inlet carton

Weight:

65 lbs. – sampler carton, 60 lbs. - inlet carton

Assembled Space Requirement:

30" round x 64" high

Power Requirement

Single phase with earth ground, 110v/60hz or 220v/50hz

Motor Specifications:

110v/60hz – Part #TE-116311, 2-stage vacuum 0.6 HP, double ball bearing thru-flow discharge, 220v/50hz – Part #TE-116312, 2-stage vacuum 0.6 HP,ball/sleeve bearing thru-flow discharge

 

Altitude Application:

to 10,000 feet above sea level

Ambient Temperature Application:

-30 to + 50ºC

Relative Humidity Application:

0 - 100%

Sample Delay Start Range:

0 - 7 days

Start up amps:

12 amps

Running amps:

4 amps

    

 

Other electrical characteristics available  

Back to Top