Enerac 3000 sem

 
Specifications  
Physical:

1. CASE: 18" x 13" x 6": Aluminum carrying case with lock.  Weight: 22 lbs.

2. PROBE: 24" L  x 3/8" OD.  Inconel probe with Hastelloy x sintered filter and 1/2" deflector mounted on permeation driver housing.  Probe housing connects to instrument via a 10 ft. Viton hose.  Max. continuous temperature: 1800 deg. F.  Max sample dew point (past dryer) 50 deg. F. @ 500 cc/min (Natural gas fuel @ 0% oxygen ).

Electrical Power:

1. Battery: 6V rechargeable, sealed lead-acid cell. Three hour continuous battery operation.  Quick 6 hour recharge. 

2. AC: 120V/60 Hz and 220V/50Hz standard.

3. DC: 11-40 VDC/3A standard

Display Power: 0.5" High by 24 Character single line LCD with backlight illumination and adjustable viewing angle.
Printer:

SEIKO 4", 40 char. per line thermal printer with line feed button and with end of paper override.  Operates in any four modes:

1. Text Mode: 25 line printout of instant values of all measured parameters.  (time req. 20 sec)

2. Plot Mode: Any one parameter vs time plotted. 3 ordinate scales: full, half, quarter.

Time scale:  Selectable, 1 sec/dot-1 min/dot in 1 sec/dot intervals.  4 parameter plot CO-OXYGEN-NOxEFFICIENCY, single scale.

3. Calibration Certification Protocal (QA/CCP):  Automatic printout of calibration test results, including sensor and filter performance status and diagnostics.

4.  External Print Mode: Prints messages sent via rs-232 port.

Storage:

1. Internal: 80 individually selectable buffers hold one complete set of measurements each in non volatile memory.  Buffer contents can be sent to printer or RS-232 port.

2. EXTERNAL(OPTIONAL):  Accessory pocket computer can store up to 300 measurements in non volatile memory. 

Communications: 1. RS-232 PORT: RS232c port (DTE), 1200 baud default, 300-9600 baud user selectable, half duplex, 1 start bit, 8 data bits, 1 stop bit, no parity
Miscellaneous

1.FUELS: 15 fuels, 3 in foreground, 12 in background standard.  Custom fuels available on request.

2. CO-ALARM:  Selectable 0-2000 PPM in 10 ppm steps.

3. Combustables in ASH: Presettable 0-100% in 5% steps.

4. Messages:  User friendly diagnostic and help messages.

5. Calibration: Autogas plus user selectable auto zero at start.

 

 

 

Operational Flexibility

 

Historically, Clean Air Permits have been established using emission factors - half of these permits may be grossly understated, thus reducing long term operational flexibility.  Many of the strategies being developed to provide future plant operational flexibility will rely on establishing accurate, defendable and cost effective emissions data.  ENERAC 3000 sem provides the most comprehensive and complete NOx measurement available.  Numerous independent studies have demonstrated ENERAC ability to supply data that meets EPA enhanced monitoring performance requirements.

 

Operator Training & Certification 

 

SEM electrochemical portable instrumentation is an important, cost-effective method to acquire compliance-level emission data. To ensure proper implementation, the operator should be trained as to the instruments capabilities.  ENERAC offers a Factory Training and Certification Program as proposed under 40 CFR 64 and detailed in the Enhanced Monitoring Reference document.

 

Remote Operation

 

2-way advanced communication and remote operation includes remote factory check and repair, and remote operation and reporting.

 

Accessories 

 

A complete line of supporting accessories is available and includes:

 

 

 

 

 

 

 

Back to Top